FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

(DRM)

2

Method demonstrates excellent repeatability and reproducibility. With all MLT data

included, repeatability ranged from 2.26-7.03% RSD and reproducibility data ranged

from 3.04-9.48% RSD.

Did Method Author Consider ERP’s Method Specific Recommendations (See web link to specific method

comments):

Any Revision to Method Scope/Applicability of Method:

Biocytin was added to the method to extend method applicability to “total” biotin

although none was detected in any of the MLT matrices.

General comments (i.e. safety) about the method:

Feedback from Users of the Methods since being awarded First Action

Official Methods

Status

One lab suggested changing the pH of the buffer to 4.0 to obtain a clear filtrate.

Method needs a detailed description of the “back Flush” technique. Syringe filters need

to be tested for possible interferences before they are used.

Additional Pros/Strengths:

Method relatively simple and easy to perform. Method has excellent repeatability and

reproducibility.

Additional Cons/Weaknesses

Method needs more detailed description of “back Flush” technique.

Supporting Information

General Comments

Performance Characteristics and compared to approved SMPR:

Analytical Range:

SRM range 0.1-150 mcg/100g. MLT sample range ~ 4-260

mcg/100g.

LOQ:

Accuracy/Recovery:

Precision (RSD

r

):

SMPR: < 6% RSD, MLT range – 2.26-7.03% with no outliers excluded.