Expert Review Panel for Infant Formula and Adult Nutrition

“Designate a Recommended Dispute Resolution Method (DRM)”

Evaluation of Method 2016-03

Title: Chloride in Infant Formula and Adult/Peidatric Nutritionals AOAC First Action Method 2016.03

MLT

Author:

Greg Jaudzems and Fengxia Zhang

Did Method Author Consider ERP’s General Recommendations:

Was the method tested using all the SPIFAN samples for precision and accuracy?

All fortified and unfortified SPIFAN matrices were tested along with IDF matrices of

cheese, butter, whey protein concentrate, whole milk powder, whole milk, and whey

powder.

Was data provided from a standard reference material?

Yes. SRM 1948a MLT mean: 701 (n=16, RSD=2.44%). NIST SRM 1849a Reference value:

7010±170 mg/kg .

Was the analysis conducted using ready-to-feed (reconstituted) infant formula?

Yes

Was rice based formula available and was it used?

No. SPIFAN kit did not include rice based formula.

Was the method tested using properly characterized standards?

Method requires silver nitrate standard meet European, British, and United States

Pharmacopoeia specifications. Participating MLT Laboratories were required to

report AgNO3 standardization data.

Is there proof of performance through system suitability?

Yes. Method requires analysis of blank and a check of the silver electrode before

beginning assay.

Is there clarity of calibration procedure?

Yes

Was the SPIFAN SLV guidelines followed?

NAP

Does the method demonstrate repeatability? To what extent?