Accuracy/Recovery:
Precision (RSD
r
):
SMPR: < 2% RSD, MLT range – 0.15-12.4% (all matrices) and 0.15-
1.68% when 2 placebo’s are excluded.
Reproducibility (RSD
R
):
SMPR: <4% RSD, MLT range – 0.95-20.8% (all matrices) and 0.95-
3.89% when 2 placebo’s are excluded.
Recommendation for Designating as Dispute Resolution Method: (Y/N)
•
Did the Author satisfactorily address ERP’s general comments?
•
Did the Author satisfactorily address ERP’s specific comments?
•
Does the method meet the approved SMPR?
Yes
•
Has the method been performing well in the laboratory?
•
Would you recommend designating this method as a dispute resolution method to the ERP?
Yes
Final Comments?
I recommend this method for final action status and consideration as a dispute resolution
method.