3

SMPR requirement (11%) and are notable given the number of labs and diversity of matrices

included in the MLT.

Precision as Reproducibility (RSDR) – Reported values range from 6.4–12.7% and are within

SMPR requirement of 15%. Reproducibility for 9 of the matrices are <10%. This is also

notable. There also is no obvious performance trend vs Product type for either Repeatability

or Reproducibility.

MLT vitamin D

3

results returned for SRM 1849a were compared to the NIST certified value.

The overall mean reported by MLT of 10.2 mcg/100g and compares favorably to the NIST

certified value of 11.1 mcg/100g.

SLV Performance Summary:

Demonstrated method repeatability, accuracy, recovery and LOQ for both vitamin D

2

and D

3

presented during original SLV meet or exceed SMPR.

Analytical Range:

Vitamin D

2

: 0.04 – 7.3 mcg/100g RTF

Vitamin D

3

: 0.06 – 11.3 mcg/100g RTF

SMPR: 0.12 – 5.1 mcg/100g RTF

LOQ:

Vitamin D

2

: 0.016 mcg/100g RTF (≤0.120 mcg/100g RTF)

Vitamin D

3

: 0.028 mcg/100g RTF (≤0.120 mcg/100g RTF)

Accuracy/Recovery:

Vitamin D

2

: 97 – 101% (mean = 98%); (SMPR 90-110%)

Vitamin D

3

: 94 – 101% (mean = 98%); (SMPR 90-110%)

vs. NIST SRM - Vitamin D3: within certified range

Precision as Repeatability (RSD

r

):

Vitamin D

2

: NAP, none of SPIFAN matrices contain vitamin D

2

Vitamin D

3

: 1.5 – 5.3% (mean = 3.0%); (SMPR ≤11%)

Recommendation:

The data produced by the MLT meet SMPR requirements for both Reproducibility and

Repeatability.