Table of Contents Table of Contents
Previous Page  61 / 65 Next Page
Information
Show Menu
Previous Page 61 / 65 Next Page
Page Background

FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Pros/Strengths:

Single day saponification and extraction

Use of UHPLC-MS/MS

Cons/Weaknesses

Only 9 collaborating labs.

Would be nice to know which instruments were used with each of the collaborating labs.

Supporting Data

General Comment:

-

Method Optimization:

Performance Characteristics:

Analytical Range:

Met SMPR

LOQ:

Met SMPR

Accuracy/Recovery:

Mean of 32 NIST 1849a results were lower than mean of range.

Precision (RSD

r

):

Met SMPR

Reproducibility (RSD

R

):

Highest RSDs were 12.4 and 12.7% which met SMPR requirement of <15%