Independent MU checks
•
Recalculates the dose, based on the plan parameters from the
planning system (or v.v.)
•
This will check (in theory)
– Amount of monitor units
– Problems with plan normalization
– Computation errors of planning system
•
Third party software
– Lots of software around (small companies)
– Check what it really checks
– Test with intentional errors