MU range checking

In house NKI development, but easy to build

Plans following a certain protocol, e.g. prostate

– Amount of MU for a VMAT plan will be similar for each patient

– Depends a little on patient size, etc

MU range check

– If patient does not fall within the range, something may have gone wrong

– Check by physics

– About 5-10%

– Usually anatomical reasons