Table of Contents Table of Contents
Previous Page  57 / 1023 Next Page
Information
Show Menu
Previous Page 57 / 1023 Next Page
Page Background

ESTRO 35 2016 S35

______________________________________________________________________________________________________

phantom, both parallel and perpendicular to the magnetic

field, and in water (waterproof chambers only). The

influence of the distribution of air around the chambers in

the SW phantom was measured by displacing the chamber in

the insert using a paper shim, approximately 1 mm thick,

positioned in different orientations between the chamber

casing and the insert.

Results:

The responses of the three waterproof chambers

measured on the MR-linac increased by 0.6% to 1.3% when

the air volume in the insert was filled with water. The

responses of the chambers on the Agility linac changed by

less than 0.3%. The angular dependence ranged from 0.9% to

2.2% in solid water on the MR-linac, but was less than 0.5% in

water on the MR-linac and less than 0.3% in SW on the Agility

linac. An example of the angular dependence of a chamber is

shown in Figure 1.

Changing the distribution of air around the chamber induced

changes of the chamber response in a magnetic field of up to

1.1%, but the change in chamber response on the Agility was

less than 0.3%.

Conclusion:

The interaction between the magnetic field and

secondary electrons in the air volume around the chamber

reduces the charge collected by between 0.6 and 1.3%. The

large angular dependence of ion chambers measured in SW in

a magnetic field appears to arise from a change of air

distribution as the chamber is moved within the insert, rather

than an intrinsic isotropy of the chamber sensitivity to

radiation. It is therefore recommended that reference

dosimetry measurements on the MR-linac be performed only

in water, rather than in SW phantoms.

OC-0076

Towards MR-Linac dosimetry: B-field effects on ion

chamber measurements in a Co-60 beam

J. Agnew

1

The Christie NHS Foundation Trust, CMPE, Manchester,

United Kingdom

1

, G. Budgell

1

, S. Duane

2

, F. O'Grady

1

, R. Young

1

2

National Physical Laboratory, Radiation Dosimetry Group,

Teddington, United Kingdom

Purpose or Objective:

To quantify the effect of small air

gaps at known positions on ionisation chamber (IC)

measurements in the presence of a strong magnetic (B-)field,

and to characterise the response of ICs over a range of B-

field strengths in the absence of air gaps.

Material and Methods:

The ratio of responses of four

commercially available ICs was measured in a Co-60 beam

with and without a 1.5T B-field (

M

1.5T

/

M

0T

) using a GMW

electromagnet unit and a 5cm pole gap. Measurements were

made in custom-built Perspex phantoms with the chamber,

beam and B-field all orthogonal. The measurements were

repeated with the phantoms at each cardinal angle (rotated

about the long axis of the ICs). The phantoms were designed

to be symmetric under rotation about this axis except for a

shallow 90° section next to the sensitive volume of the ICs.

The measurements were repeated with the air gap removed

by introducing water to the phantom cavity. For the PTW

30013 chamber further measurements were performed after

introducing a small (approximately 30 mm

3

) bubble into the

recess when the cavity was otherwise filled with water,

which was made possible by the novel phantom design. The

measurements in water were repeated with additional build-

up material and in multiple phantoms at a single phantom

orientation.

Measurements were also taken to characterise the ratio of

responses for five ICs over a range of B-field strengths (0 – 2T

in 0.25T increments).

Results:

For all 4 ICs in the rotating setup, the response

varied consistently with the position of the recess when the

air gap was present, with the lowest value of

M

1.5T

/

M

0T

obtained when the recess was upstream of the IC. The

maximum peak-to-peak (PTP) variation was 8.8%, obtained

for the PTW 31006 ‘Pinpoint’ IC, and the minimum was 1.1%,

obtained for the Exradin A1SL IC. This variation all but

disappeared (maximum PTP variation 0.7%, seen for PTW

31010 IC) when the air gap was removed. A large (3.9%) PTP

variation was observed for the PTW 30013 when an air bubble

was inserted into an otherwise airless setup (0.2% variation

without air gap).

Conclusion:

Small air gaps are responsible for large variations

in IC response in the presence of a magnetic field. These

variations can be eliminated by introducing water into the

cavity, but even small bubbles will cause large variations in

the response. Further, IC response in the presence of a 1.5T

B-field is insensitive to changes in depth and scatter

conditions of the phantoms investigated here. Each IC has

different

M

/

M

0T

response across the range of B-field strength

0 – 2T.

OC-0077

Dual energy CT proton stopping power ratio calibration:

Validation with animal tissues

Y. Xie

1

University of Pennsylvania, Department of Radiation

Oncology, Philadelphia, USA

1

, L. Yin

1

, C. Ainsley

1

, J. McDonough

1

, T. Solberg

1

, A.

Lin

1

, B.K. Teo

1

Purpose or Objective:

One main source of uncertainty in

proton therapy is the conversion of Hounsfield Unit (HU) to