D. By Means of U-Tube Oscillation – Digital Density Meter Method

(a)

Direct.

– Specific Gravity of juices.

(b)

Principal

Natural frequency of hollow oscillator varies with density of

material that fills tube. Electronic measurement of a time period is

converted to digital display of density, specific gravity and Brix.

(c)

Apparatus and Reagents

(1)

Digital density meter

: Anton Paar® GmbH, Anton-Paar-

Str.20, A-8054 Graz, Austria – Europe;

(www.anton-paar.com)

.

DMA 4500M digital analyzer with U-shaped, oscillating sample tube.

(2)

Hypodermic syringe

. As recommended in density meter

instructions.

(d)

Calibration of Density Meter

Calibrate digital density meter at 20.0°±0.1°C with air and H

2

O.

Use hypodermic syringe to inject liquids into analysis tube. Fill tube

completely so no gas or solids are entrapped

(e)

Determination

Inject test sample, if carbonated beverage (ensure

decarbonation), into filling tube. Be sure no gas is entrapped in tube.

Let thermostated analysis tube reach thermal equilibrium. Obtain

specific gravity measurement if desired result are Brix then convert to

Brix using

AOAC942.33

– Degrees Brix, specific gravity, and

degrees Baume of sugar solutions (Plato Table).

References: Anton Paar® GmbH.

J. of AOAC Intl

, Vol XXX, No.

XX (20XX).

J. Am. Soc. Brew. Chem

.

36

, 118 (1978).

Methods of

Analysis

(1992) 8

th

Ed., American Society of Brewing Chemists,

3340 Pilot Knob Rd, St. Paul, MN 55121, USA, Beer 2-B.

AOAC Research Institute

Expert Review Panel Use Only