Table of Contents Table of Contents
Previous Page  373 / 435 Next Page
Information
Show Menu
Previous Page 373 / 435 Next Page
Page Background

Multi Laboratory Test – AOAC First Action Method 2012.22 Ascorbic Acid (Vitamin C) in Infant Formula and

Adult/Pediatric Nutritionals.

RDLS-RD150006

E. Campos Giménez (NRC, Lausanne)

Page 11

CONFIDENTIAL

The contents of this document constitute a trade secret and the intellectual property rights thereto belong to Nestlé S. A., Vevey, Switzerland.

This document may not be reproduced or disclosed to third parties without prior authorisation.

Part 2. Multi Laboratory Test,

Continued

Results

(continued)

In general, the precision results (repeatability and reproducibility) are well within

the limits stated in the SMPR. Repeatability ranged from 1.4 to 7.3 %, in

accordance with data obtained during SLV

1

, with reported RSD

r

from 0.5 to

2.5 %. Meanwhile reproducibility ranged from 3.7 to 11.4 %. Finally the HorRat

values are all below 2, from 0.45 to 1.54. Except for the two samples already

discussed above.

1

Internal Nestlé document RDLS-RD130018

;

Single Laboratory Validation – AOAC First Action Method

2012.22 Vitamin C in Infant Formula and Adult/Pediatric Nutritional Formula by UPLC-UV

; E. Campos Giménez;

February 14

th

2013.

AOAC OMB Meeting Book

345