1.
Injection volume.- 5µL.
77
2.
Autosampler temperature.- 10 °C.
78
3.
Column temperature.- 25 °C.
79
4.
Flow rate.- 0.35mL/min
80
5.
Run time.- 4.0 minutes
81
6. Mobile phase.- Sodium acetate (50 mmol/L, pH 5.4), decylamine (1.6g),
82
acetonitrile (1 %), TCEP (40 mg/L).
83
NOTE:
At the end of each analytical series, rinse column with acetonitrile-water (1+1, v/v)
84
for 10 minutes at 0.4 ml/min
.
85
(b).
System suitability test
.- Equilibrate the chromatographic system for ≥0.5 h.
86
Inject a working standard solution of ascorbic acid at least 6 times and check
87
peak retention times and response (peak height or area). Inject working
88
standard solutions on a regular basis within a series of analyses. Ensure orotic
89
acid and isoascorbic acid are fully resolved from ascorbic acid by injecting
90
separated standard solutions of both compounds (prepared as stated for
91
ascorbic acid). If not resolved, the pH of the mobile phase may be decreased to
92
5.0, or the amount of acetonitrile increased.
93
(c).
Calibration
.- Make single injections of working standard solutions at least at
94
the beginning and the end of each analytical series. Establish the calibration
95
curve (seven points) by plotting peak response (height or area) versus ascorbic
96
acid concentration. Perform linear regression. Calculate the slope (S) and
97
intercept (I) of the calibration curve.
98
(d).
Analysis
.- Make single injections of sample solutions.
99
AOAC 2012.22 (FINAL VERSION REVIEW) - JUNE 2016