1.

Injection volume.- 5µL.

77

2.

Autosampler temperature.- 10 °C.

78

3.

Column temperature.- 25 °C.

79

4.

Flow rate.- 0.35mL/min

80

5.

Run time.- 4.0 minutes

81

6. Mobile phase.- Sodium acetate (50 mmol/L, pH 5.4), decylamine (1.6g),

82

acetonitrile (1 %), TCEP (40 mg/L).

83

NOTE:

At the end of each analytical series, rinse column with acetonitrile-water (1+1, v/v)

84

for 10 minutes at 0.4 ml/min

.

85

(b).

System suitability test

.- Equilibrate the chromatographic system for ≥0.5 h.

86

Inject a working standard solution of ascorbic acid at least 6 times and check

87

peak retention times and response (peak height or area). Inject working

88

standard solutions on a regular basis within a series of analyses. Ensure orotic

89

acid and isoascorbic acid are fully resolved from ascorbic acid by injecting

90

separated standard solutions of both compounds (prepared as stated for

91

ascorbic acid). If not resolved, the pH of the mobile phase may be decreased to

92

5.0, or the amount of acetonitrile increased.

93

(c).

Calibration

.- Make single injections of working standard solutions at least at

94

the beginning and the end of each analytical series. Establish the calibration

95

curve (seven points) by plotting peak response (height or area) versus ascorbic

96

acid concentration. Perform linear regression. Calculate the slope (S) and

97

intercept (I) of the calibration curve.

98

(d).

Analysis

.- Make single injections of sample solutions.

99

AOAC 2012.22 (FINAL VERSION REVIEW) - JUNE 2016

FOR ERP USE ONLY

DO NOT DISTRIBUTE

AOAC OMB Meeting Book

393