14. METHOD COMPARISON RESULTS
309
The results of the method comparison studies are summarized in Tables 2 and 3 below. For
310
all sample types and enrichment methods tested, the updated BAX® System method
311
returned results that were identical to the certified method. POD analysis demonstrates no
312
statistically significant difference between results of the updated BAX® System method and
313
the reference culture methods.
314
315
Fluorescent Probe Performance
316
Performance of the fluorescent probes in the additional eight cycles do not appear to suffer
317
either by photo bleaching, which would lead to decreased signal for a positive reaction, or by
318
quencher degradation, which would lead to a false positive rise in signal. Thousands of reactions
319
have been run internally to date in the algorithm training and validation phases of the modified
320
protocol. Examples of typical reactions are shown below in Figure 1 (negative reactions) and
321
Figure 2 (positive reactions). Data in these figures is typical of data obtained during algorithm
322
training. Additional data is available upon request.
323
Dupont BAX Salmonella PTM Report
Modification Approved 2012 / PTM Certification No. 081201
For Expert Review Panel Use Only
Do Not Distribute