Acuros - Varian

Transport of particles in medium

Electron fluence calculated in each voxel – differential in energy

Mass stopping power applied either as tissue or medium when dose is

determined

o

The default method used to determine the material composition of a

given voxel in a 3D image is based on the HU value.

o

In the only commercial implementation of the GBBS, this is converted

to

mass density

using the CT scanner calibration curve. For voxels with

density <3 g/cm

3

, only a biological material can be assigned: lung,

adipose tissue, muscle, cartilage, or bone.

o

The material corresponding to a given HU is hard-coded into a lookup

table..

Thus or (user select)