Phantoms for TPS verification and plan QA: use of solid plastics

If the solid plastic material is not water equivalent in the kV beam energy region at

which the CT scan is being performed, then the CT calibration curve will not correctly

represent the relationship between relative density and Hounsfield numbers.

In such a case, it is best to override the material density with the density of the

medium as given by its manufacturer

Use this description of the phantom for dose calculations and comparisons

of these against measurement in the phantom