Phantoms for TPS verification and plan QA: use of solid plastics

To compare measured doses in

plastic phantoms

against doses calculated by

TPSs, the phantoms needs to be corrected represented in the TPS

For some TPSs, the HN to relative mass density calibration does not include

an assignment of HN to the density and composition of plastic materials,

because plastics area not included in the list of available materials

In such cases, to avoid systematic deviations in the dosimetric validation of

dose engines, an

effective mass density

is being assigned to the plastic

medium.

An implementation of such a workaround can be found in Oncentra.

An effective mass density as implemented in the RayStation, is

computed for all media and aims to account for any systematic errors

introduced in the assignment of mass density to CT-HN as a result of photon

attenuation in the kV energy range (CT energies)

HN to relative mass density calibration

Oncentra