Novalis ExacTrac

Patient plan was prepared for morning staff

Treatment parameters transferred to both data bases

(Varis and ExacTrac: labeled “ready for approval”) for

QA purposes

CTV was rejected and adjusted at morning staff

This resulted in a change of isocentre co-ordinates

Final plan was transferred to Varis R&V (labeled

“approved by staff dd/mm/yy) … but not to ExacTrac

RTT discovered discrepancy in isocentre co-ordinates

while dubble checking print out of treatment chart.