Mandatory Recommended Optional
Fluoroscopy at simulation for evaluation of tumor motion
0
0
6
Abdominal compression system
0
0
5
Active breathing coordinator system (e.g. ABC system)
0
2
5
Respiration correlated 4D-PET-CT
0
0
8
Implantable fiducial marker system
0
1
6
Implantable transponders e.g. Calypso System
0
0
7
Audio and / or visual breathing motion monitoring system
for breathing feedback
0
2
6
Surface Scanner
0
1
5
External breathing motion monitoring system in the
treatment room (e.g. RPM system)
0
3
5
Linac with gated beam delivery mode
0
2
6
Flattening filter free (FFF) delivery mode
0
2
6
Very high resolution MLC < 5mm
0
2
6
Robotic 6 degrees of freedom (DOF) treatment couch
1
2
5
Additional technologies
•
Most additional technologies optional
07.09.17
ESTRO SBRT COURSE 2017
7
/
Staffing and Credentialing
Mandatory
Written departmental protocol
covering all
mandatory aspects of SBRT practice
8
Site-specific SBRT implementation &
application based on
a multi-disciplinary
project team
involving Clinicians, Physicists &
RTTs
8
Structured follow-up
and assessment of
clinical outcomes (e.g. local control, toxicity)
8
•
Mandatory: Protocols, multi-professional team &
structured follow-up
07.09.17
ESTRO SBRT COURSE 2017
8