A Practical Example: SRT Case

Intracranial lesion: 3 x 8 Gy @ 80%

N=3,

b

=0.84

SD of the penumbra is 3.2 mm

s

pen

=3.2 mm

E2E test device error (

S

) = 0.4 mm

S

1

=0.4 mm

Localization (delineation) error = 1.0 mm (1 SD)

S

2

=1.0 mm

Systematic error = 0.5 mm (1 SD)

S

eff

=0.58 mm

Random error = 0.5 mm (1 SD)

s

eff

=0.41 mm

Intra-fraction error = 0.5 mm ( 1 SD)

s

eff

=0.20 mm