MRI to CT Registration
Keep patient positioning the same for MRI and CT scanning
▪
Flat table top
▪
Similar bladder filling and rectum filling instructions (also for treatment)
▪
No rectal coil!!!!
Markers are poorly visible on standard MRI sequences that are used to
visualize the tumor
▪
Use additional sequence to visualize markers in order to facilitate MRI-to-CT
registration
Calypso markers give large artefact in MRI
▪
Do MRI before implantation of markers
Discuss with the radiologist the MRI settings and sequences
▪
A MRI for radiotherapy has other requirements as for radiology purposes