Spine SBRT: Required accuracy
maximum tolerable errors on average :
1 mm (transversal plane)
4 mm (SI direction)
3.5
°
(spinal cord dose within
±
5% of prescribed dose)
Simulated transversal patient set-up errors (0.5–10 mm)
Guckenberger R&O 84, 2007 p56