An example

Description of the treatment planning system

Ø

The treatment planning was made with Multiplan

version 2.2.0, the treatment planning system of

the Cyberknife (Accuray, Sunnyvale, CA).

Multiplan has implemented both equivalent path

length (EPL) and Monte Carlo (MC) dose

calculation algorithms. The MC dose calculation

algorithm was validated (Grofsmid et al., 2010).

The treatment plans in Multiplan were first

recalculated with EPL, and then with MC. This

was done to eliminate subtle differences

between the OnTarget and Multiplan treatment

planning systems. A high-resolution grid (256 ×

256) was used for the EPL and MC calculations

and the type-A uncertainty in the MC calculation

was set to 2 %. MC computation time was

approximately 5–10 minutes

SBRT 2017 - D. Verellen