Questions from the audience

Some careful considerations, some answers

Ø

Prescribing to an enveloping isodose line is fine, but what isodose

should we use? Isodose compared to what normalization point?

§

Isodose compared to isocentre dose or D

near-max

?

§

Which isodose 60%-80%, 95%?

§

Unfortunately ICRU only says “an enveloping isodose” must be used, so:

“yes it is still confusing”.

§

Fortunately, with what should be reported (D

50%

to PTV and GTV, and

D

near-max

and D

near-min

), you should be able to translate a published result

to your prescrition method.

o

eg, if you use 15 Gy to 60% of isocentre dose, it will become difficult to

compare that to a study reporting D

50%

= 15Gy, D

near-max

= 17Gy

Ø

Very confusing in the ICRU 91 is the concept in that the accepted

plan becomes the “prescription” for reporting, this is contradicting

daily practice

SBRT 2017 - D. Verellen