Table of Contents Table of Contents
Previous Page  191 / 1096 Next Page
Information
Show Menu
Previous Page 191 / 1096 Next Page
Page Background

S178

ESTRO 36

_______________________________________________________________________________________________

1

Université Catholique de Louvain- Institute of

Experimental & Clinical Research, Molecular Imaging-

Radiotherapy & Oncology, Brussels, Belgium

2

Centre Antoine Lacassagne, Medical Physics, Nice,

France

3

EBG MedAustron GmbH, Medical Physics, Wiener

Neustadt, Austria

4

National Physical Laboratory, Acoustics and Ionising

Radiation Division, Teddington, United Kingdom

5

Cliniques Universitaire St-Luc, Radiotherapy and

Oncology Dep., Brussels, Belgium

6

IBA Dosimetry GmbH, Schwarzenbruck, Germany

Purpose or Objective

The main application of calorimeters in standards

laboratories is as primary standard of absorbed dose to

water against which ionisation chambers (ICs) are

calibrated. At present, no calorimeter is established as a

primary standard instrument in proton beams.

Based on the absorbed dose-formalism of IAEA TRS-398,

this work describes a direct comparison between a water

calorimeter (WCal) and plane-parallel ICs in clinical pulsed

pencil beam scanning (PBS) proton beams, delivered by a

synchrocyclotron. The temporal beam characteristics and

the absence of a dosimetry protocol for such beams create

significant challenges in absorbed dose determination.

The aim of this work is to demonstrate the feasibility a

water calorimetry in pulsed PBS beams.

Material and Methods

The method consisted in comparing the response of WCal

and ICs (PPC40 and PPC05) in the same reference

conditions. Measurements have been performed at a

depth of 3.1 cm using two mono-layers maps of proton

beams (10 x 10 cm²), with incident beam energies of 96.17

MeV (range in water = 6.8 g/cm²) and 226.08 MeV (range

in water = 31.7 g/cm²), respectively. The response of the

WCal is corrected for heat transfer (calculated using

numerical simulations based on finite element method)

and non-water material inside the WCal (using

experimentally derived factors). Using hydrogen-

saturated high-purity water in the WCal, the chemical

heat defect is assumed to be zero. Classical correction

factors are applied to the response of ICs: temperature

and pressure, polarity and recombination (k

s

). k

s

was

studied in detail due to the very high beam dose rate used

with the delivery method.

Results

Table 1 shows preliminary relative differences of D

w

measured with WCal and IC, during two independent

experimental campaigns, for both energies. A small

positioning uncertainty could explain that the ratios

obtained during campaign B are higher for the low energy

beam. For campaign A, however, ratios are higher for the

high energy beam, which cannot be explained by a

positioning uncertainty. A new campaign is planned to

repeat the measurement of correction factors to improve

the statistics of the results.

Conclusion

The preliminary results are very encouraging and

demonstrate that water calorimetry is feasible in a clinical

pulsed PBS proton beam. The absolute relative differences

between D

w

derived from WCal and IC are inferior to 2%,

which is within the tolerance of the IAEA TRS-398 protocol.

Due to the depth-dose distribution, a depth inferior to 3.1

cm (e.g. 2 cm where the gradient is lower) would be more

suitable to minimise the uncertainty in positioning.

Further numerical and experimental investigations are

planned to confirm and consolidate correction factors and

determine the overall uncertainty on absorbed dose-to-

water obtained using each system. The next experimental

step is to perform the same experimental comparison for

a real clinical situation: a dose cube of 10 x 10 x 10 cm³,

created by a superposition of mono-energetic layers.

OC-0340 Validation of HU to mass density conversion

curve: Proton range measurements in animal tissues

J. Góra

1

, G. Kragl

1

, S. Vatnitsky

1

, T. Böhlen

1

, M.

Teichmeister

1

, M. Stock

1

1

EBG MedAustron GmbH, Medical Physics, Wiener

Neustadt, Austria

Purpose or Objective

Proton dose calculation in the treatment planning system

(TPS) is based on HU information taken from the CT scans

and its relation to the relative stopping powers (RSP).

However, tissue equivalent substitutes commonly used in

the process of conversion curve definition may not reflect

precisely the properties of real, human tissues. Therefore,

various animal tissues were used for validation of the CT

number to mass density (MD) conversion curves

implemented in the TPS (RayStation v5.0.2).

Material and Methods

10 animal tissue samples (pig) were used in this study

(muscle, brain, bone, blood, liver, spleen, lung, fat,

kidney and heart). Each sample was prepared and

wrapped separately. 3-4 tissues were placed in dedicated

phantoms (head and pelvis) at a time and CT scans were

taken in the clinically accepted planning protocols.

Specially designed PMMA phantoms where composed of

two parts: a) an internal box, which could fit the animal

tissues inside, b) the outer PMMA cover, designed to

simulate pelvis (see fig.1c) and head during CT scan. The

design of the phantoms not only helped to reduce imaging

artefacts but also allowed to apply a slight pressure on the

tissues in order to remove unwanted air. Subsequently,

the tissue phantom was attached to the front of the water

phantom, where with the use of 2 Bragg peak chambers,

range measurements were performed. All measurements

were performed within 24h after the animal was

slaughtered with the use of one, central, 160.3 MeV pencil

beam. For each sample, multiple irradiation positions

were chosen in a very precise matter, as it was extremely

important to choose the most homogeneous path through

which the proton beam would pass. Acquired CT data was

used to read out the HU, correlate them with the

measured RSP and validate against implemented CT

number to MD conversion curves.

Results

Figure 1, shows the comparison between measured RSP

and HU for real tissue samples and implemented

conversion curve in the TPS a), CT scan of the adult,

abdomen protocol b), and measurement set-up c). The

measured data for all soft tissues were found to be within

1% agreement with the calculated data. Only for lung

tissue the deviations were up to 3.5%. For bone, both the

difficulty in assessing the actual thickness of the part

where the beam was passing through, as well as the

inhomogeneous nature of this tissue, prevented us from

the accurate RSP assessment. However, for 2

measurements out of 3, the measured RSP where within

3.5% uncertainty.

Conclusion

The experimental validation of the conversion curve

resulted in good agreement between measured and

calculated data, therefore we can use it in the clinical set-

up with confidence. There is a number of uncertainty

sources related to these measurements, starting from HU

to RSP model, real tissue heterogeneities or uncertainties

related to acquisition of the CT data due to beam

hardening. The last one, we tried to minimize by using

especially dedicated phantom.