Table of Contents Table of Contents
Previous Page  274 / 1096 Next Page
Information
Show Menu
Previous Page 274 / 1096 Next Page
Page Background

S261

ESTRO 36

_______________________________________________________________________________________________

contour OAR located within 3cm from PTV

opt

, based on the

daily MR imaging (SMART

3CM

). Optimization structures are

automatically adapted to the new anatomy, and re-

optimization is performed using exactly the same plan

parameters. This limited re-contouring strategy was

evaluated by comparing 45 previously delivered fractions

against a simulated standard (re-)planning method using

full-scale OAR (re-)contouring, where optimization

objectives were used for the whole organs (SMART

FULLOAR

).

Baseline plans for OAR were created that had identical

plan quality as were achieved for SMART

3CM

. Efficiency of

both strategies was scored according to the number of

optimizations needed to generate a high quality plan. Plan

quality was assessed using PTV coverage (V95%) and

institutional OAR constraints (V33Gy and V25Gy).

Results

The SMART

3CM

baseline plans required a lower number of

optimizations than SMART

FULLOAR

(4 vs 17 on average).

PTV

OPT

coverage with both strategies was identical in all

fractions (median V95%=93±6.8%). SMART

3CM

uniformly

resulted in plans which complied with the V33Gy dose

constraint for OARs, whereas SMART

FULLOAR

failed in 35% of

the cases to adhere to the V33Gy dose constraint

according to the clinical protocol. Both strategies

achieved V25Gy values lower than 20 cc for all OARs in

every fraction. However, on average, SMART

3CM

resulted in

a

lower

V25Gy

than

SMART

FULLOAR

(Fig.2).

Conclusion

This fast and robust (re-)planning approach for SBRT to

pancreatic tumors requires clinicians to only re-contour

OARs located within 3cm of the PTV

OPT.

Spatially

partitioned optimization structures within this 3 cm region

allowed for optimal OAR sparing, and adequate target

coverage, using exactly the same plan parameters.

OC-0491 Quality assurance of a novel table mounted

imaging device integrated in a patient positioning

system

A. Utz

1

, A. Ableitinger

1

, A. Zechner

1

, M. Mumot

1

, M.

Teichmeister

1

, P. Steininger

2

, H. Deutschmann

2

, M.

Stock

1

1

EBG MedAustron GmbH, Medical Physics, Wiener

Neustadt, Austria

2

medPhoton GmbH, Medical Physics, Salzburg, Austria

Purpose or Objective

Image guided radiation therapy (IGRT) aims to reduce

margins and subsequently increase dose sparing for OAR.

The majority of image guidance procedures are based on

ceiling/floor- or gantry mounted imaging devices. In our

particle therapy center a novel approach for patient

alignment was introduced. The imaging system (imaging

ring) is mounted on the treatment table and as such,

allows high imaging flexibility e.g. CBCT or planar imaging

at different table positions. The goal was to establish a

phantom and a concept for a quality assurance procedure

for the whole IGRT workflow.

Material and Methods

The IGRT Phantom consists of a PMMA cube with steel

fiducials, which can be placed in predefined offset

positions on a baseplate to simulate clinical patient shifts.

An additional support structure is used to lift the cube.

Holes on the upper corners of the cube allow to

independently determine the absolute position with a

lasertracker (see figure 1a). A CT imageset of the cube in

a reference position serves as planning CT. The baseplate

is indexed on the patient couch. The cube can be moved

to a predefined translational and rotational offset position

on the couch. Two planar images were acquired and

registered to the CT.The calculated correction vector was

applied by the patient positioning system. This workflow

was repeated at three index positions (equal distributed

along the treatment volume), payloads (0kg, 100kg,

150kg), cube offsets (long.: 10mm, lat.: 15mm, vert.: -