Table of Contents Table of Contents
Previous Page  816 / 1096 Next Page
Information
Show Menu
Previous Page 816 / 1096 Next Page
Page Background

S800

ESTRO 36

_______________________________________________________________________________________________

Material and Methods

In the treatment prostate plan with VMAT high-

fractionated (2x7.5Gy), FFF technique the errors of dose

(differences ±1%; 2%; 3%; 5% 7%, 10%), collimator angle

(rotations in both directions: 0.5; 1.0; 1.5; 2.0; 2.5; 3.0)

and MLC shifts were introduced. For each modified plan,

the pre-treatment verification plan was created and

measured with 2D-arrays: 729 and SRS 1000 with rotational

phantom Octavius® 4D and Verisoft 6.1 software with DVH

option (PTW, Freiburg, Germany). Measured (with errors)

and calculated (reference plan) dose distributions were

analyzed with 3D gamma evaluation method for various

tolerance parameters DTA [mm] and DD [%] 1.0; 1.5; 2.0;

2.5; 3.0, by global and local dose methods with a 5%

threshold. To detect errors, the achieved score should be

less than the assumed tolerance of 95%. Additional the

DVHs from error-induced and reference plan were

analyzed for CTV D

50

, D

98,

D

2

, and D

25

, D

50

for OARs.

Results

For 12 error-induced plan with dose discrepancies, proper

detection for 729 and SRS 1000 were obtained as follows:

3/12 and 6/12 (G3%/3mm); 8/12 and 6/12 (L3%/3mm);

8/12 and 7/12 (G2%/2mm); 8/12 and 8/12 (L2%/2mm).

The rotations of collimator were detected >3° for 729 and

>2° for SRS 1000. The MLC errors were discovered for plans

with 1 leaf (MLC1) and 1 pair of leaves (MLC2) blocked, for

all leaves shifted about 0.05cm (MLC3) misalignment

weren’t indicated so obvious. The clinical relevance of

plan with MLC errors and chosen discrepancies for

collimator rotation (3°) and dose differences (+5%) were

presented in the table 1.

Conclusion

To more sophisticated analysis the gamma criteria should

be less than 3%/3mm or/and local dose method should be

used. The resolution of used detector is crucial and should

be high for better interpretation of results. Gamma

method presents some statistic data, for scrutiny analysis

the clinical interpretation should be assessed.

EP-1493 Machine record parameters or Epid based

data for ART QA. A comparison of two scenarios.

P. Haering

1

, C. Lang

1

, M. Splinter

1

1

DKFZ, E040, heidelberg, Germany

Purpose or Objective

Using machine record files and Epid based dosimetry is

popular for machine and patient related QA, as this may

also work for adaptive treatment approaches. The

Siemens Artiste treatment machine used here, allows a

comparison of both methods in one session. Exit images

and all relevant machine parameters are included in the

image header collected during treatment. Here we

present results of a comparison between QA dose

recalculations based on the two sources, exit images and

machine recorded parameters.

Material and Methods

A software tool was developed that allows for the

extraction of the relevant parameters (MLC-positions, MU,

etc.) from the machine records as well as from the Epid

measured exit fluencies. While machine data had to

undergo a reformat to be used for recalculation, the exit

fluencies need more attention. Here both, the delivered

fluence as well as the absorption in the

patient do play a

role. Therefore both have to be separated to receive

reliable MLC positions. The algorithm used first generates

an image containing only absorption information for the

beam using this to remove this influence on the MLC

positions. MUs were used from the parameter file, as the

fluence uncertainties on the EPID images have shown to

be to large to be used for that purpose. The extracted

parameters are then inserted in a newly generated Dicom

RT-Plan file that then can be used in the treatment

planning system (here Raystation, Raysearch) to

recalculate the dose. Dose distributions (Epid based,

parameter file based and originally planned) are then

compared.

Results

Measuring exit doses with the EPID was a simple task and

could be done for all coplanar field sets. The software tool

made it simple to extract all the needed parameter from

the files and images resulting in 2 new Dicom plan files.

Dose recalculation was done by just importing the new

plan files to Raystation. Comparing the original dose

distribution to the machine file based one showed almost

no difference at all (< 0.7%), as MU and leaf position

differences where quite small. This might also be

grounded in the used calculation grid of 2mm size. MLC

positions derived from EPID images show much larger

differences. Here detection uncertainties, EPID

positioning and the resulting image resolution of 0.3mm

do play a major role. This resulted in in noticeable

differences in the dose gradients regions. Absolute dose

differences where below 1.5%.

Conclusion

Recalculating doses based on EPID and machine based

parameters is a possible way for QA in an adaptive

treatment approach. As QA parameters are taken from

information that is given anyway or that can be easily

generated, it does not complicate the procedure of

frequent replanning. Results are as expected quite good

for the machine file approach while higher discrepancies

were found using EPID data. Main problem we face here is

that especially for the machine file based version we do

not have full independent data sources.

EP-1494 The MedAustron proton gantry: nozzle design

recommendations based on Monte Carlo simulations

H. Fuchs

1,2

, L. Grevillot

2

, A. Elia

2

, A. Carlino

2,3

, J.

Osorio

2

, V. Letellier

2

, R. Dreindl

2

, M. Stock

2

, S. Vatnitsky

2

1

Medizinische Universität Wien Medical University of

Vienna, Department of Radiation Oncology & Christian

Doppler Laboratory for Medical Radiation Research for

Radiation Oncology, Vienna, Austria

2

MedAustron lon Therapy Center, Department of Medical

Physics, Wr. Neustadt, Austria

3

University of Palermo, Department of Physics and

Chemistry, Palermo, Italy

Purpose or Objective

MedAustron is equipped with one vertical and three

horizontal fixed beam lines and one proton gantry based

on the PSI gantry 2 design for patient treatments. This

work focuses on simulations and design considerations for

the proton gantry nozzle, allowing an optimization of

beam delivery properties at isocenter.

Material and Methods

Different gantry nozzle designs were evaluated using

Gate/Geant4 Monte Carlo (MC) simulations: air filled

nozzle, helium filled nozzle, full vacuum nozzle, moving

snout, compacting of nozzle elements (vacuum window

and monitors). Design considerations were based on the