Head scatter: equivalent square of a rectangular field

Linac head

Determination of weighting factor G:

a. Experimentally: set of Sc values measured for

square and rectangular fields (e.g. keeping one

collimator setting fixed and varying the other

one)

b. From treatment head geometry: relative weight

(

G

) depends on distance from the X-ray

source

primary collimator

wedge

monitor chamber

upper Y-jaws

lower X-jaws

flattening filter

MLC leaf

S

f

S

U

S

L

S

*

S

UI

S

LI

G = (S

L

/ S

U

)

2

Yu and Sloboda (1995)

isocentre

G = (S

L

/ S

LI

) (S

UI

/ S

U

)

Kim et al (1997)

Typical values: 1.4 – 2.0