• The signal from the monitor chamber can be affected from particles which have

backscattered from the upper part of the jaws into the chamber (dependent on linac head

and monitor chamber design, )

Backscatter into the monitor chamber

• At narrower collimations there is more backscatter than in larger fields

monitor

chamber reaches faster its pre-set value

linac relative output decreases with

decreasing field size

Verhaegen & Seuntjens, Phys. Med. Biol, 48 (2003) R107-R164