E. coli Detection Workflows 8/2015

38

Comments and suggestions

Weak or no signal in the internal amplification control

a) The PCR conditions do

not comply with the

protocol

Check that PCR conditions match the cycling

protocols in Table 8, page 25. Repeat PCR with

corrected settings, if necessary.

b) The PCR was inhibited Use the recommended DNA isolation method in

this workflow. If there is inhibition, dilute DNA

sample and repeat PCR

c) The storage conditions

for one or more kit

components did not

comply with the

instructions given in

“Storage” (page 7)

Check the storage conditions and the expiration

date (see the kit label) of the reagents and use a

new kit, if necessary.

d) The

mericon

PCR Assay

has expired

Check the storage conditions and the expiration

date (see the kit label) of the reagents and use a

new kit, if necessary.

Signals present for the negative controls

a) Contamination

occurred during PCR

setup

Repeat the PCR with new reagents in replicates.

If possible, close the PCR tubes directly after

addition of the sample to be tested.

Make sure to pipet the positive controls last.

Make sure that the work space and instruments

are decontaminated at regular intervals.

b) Contamination

occurred during

extraction

Repeat the extraction and PCR of the sample to

be tested using new reagents.

Make sure that the work space and instruments

are decontaminated at regular intervals.