Independent MU checks

Recalculates the dose, based on the plan parameters from the

planning system (or v.v.)

This will check (in theory)

– Amount of monitor units

– Problems with plan normalization

– Computation errors of planning system

Third party software

– Lots of software around (small companies)

– Check what it really checks

– Test with intentional errors