E. Apparatus and Conditions

(

a

)

GC-MS analysis.—

(

1

)

GC-MS system

.—Model 7890A gas chromatograph connected to a Model 5975C

mass selective detector with electron ionization (EI) source and equipped with a Model 7683

autosampler and Chemstation data processing software system (Agilent Technologies, Wilmington, DE,

USA), or equivalent.

(

2

)

Column

.--DB-1701 capillary column (30 m × 0.25 mm × 0.25 μm), or equivalent.

(

3

)

Column temperature

.—40°C hold 1 min, at 30°C/min to 130°C, at 5°C/min to 250°C, at 10°C/min to

300°C, hold 5 min.

(

4

)

Carrier gas

.--Helium, purity ≥99.999%, flow rate 1.2 mL/min.

(

5

)

Injection port temperature

.--290°C.

(

6

)

Injection volume

.--1 μL.

(

7

)

Injection mode

.--Splitless, purge on after 1.5 min.

(

8

)

Ionization mode

.—EI.

(

9

)

Ion source polarity

.--Positive ion.

(

10

)

Ionization voltage

.--70 eV.

(

11

)

Ion source temperature

.--230°C.

(

12

)

GC-MS interface temperature

.--280°C.

(

13

)

Solvent delay

.--14 min.

(

14

)

Ion monitoring mode

.--Selected ion monitoring (SIM), one quantifying ion and two qualifying ions

are selected for each compound. The retention times, quantifying ions, qualifying ions, and the expected

ion abundances for each of the 20 pesticides included in the study and heptachlor epoxide are listed in

Table

2014.09A

. SIM acquisition parameters for ions monitored by GC-MS are shown in Table

2014.09B

.

(

b

)

GC-MS/MS analysis.--

(

1

)

GC-MS/MS system

.—Model 7890A gas chromatograph connected to a

Model 7000B triple quadrupole mass spectrometer with EI source and equipped with a Model 7693

autosampler and Mass Hunter data processing software system (Agilent Technologies), or equivalent.

(

2

) Operating conditions are the same as for GC-MS with the exception that the ion monitoring mode is

by selected reaction monitoring (SRM) and monitoring one precursor ion and two product ion

transitions.

(

3

) The monitored ion transitions and the collision energies for the 20 pesticides of interest in the study

and heptachlor epoxide are shown in Table

2014.09C

. The SRM acquisition partameters for the precursor