20

Correcting Patient position

How stable is a mask?

Study

Immobilisation system

Imaging modality

Position error

Boda Heggeman 2006

Thermoplastic mask

CBCT

1.8mm

±

0.7mm1.3mm

Scotsch cast mask

1.3mm

±

1.4mm

Masi 2008

Thermoplasic mask & Bite

block

CBCT

<1.0mm

Bite block

<1.0mm

Lamda 2009

BrainLab mask

2D kV images

0.5mm

±

0.3mm

Ramakrishna 2010

BrainLab mask

2D kV images

0.7mm

±

0.5mm

Guckenberger 2007

Scotsch cast mask

CBCT

0.8mm

±

0.4mm

Thermoplastic mask

0.8mm

±

0.5mm

Courtesy M. Guckenberger, ESTRO IGRT course