Novalis ExacTrac

Patient plan was prepared for morning staff

Treatment parameters transferred to both data bases (Varis and ExacTrac: labeled

“ready for approval”) for QA purposes

CTV was rejected and adjusted at morning staff

This resulted in a change of isocentre co-ordinates

Final plan was transferred to Varis R&V (labeled “approved by staff dd/mm/yy) … but not

to ExacTrac

RTT discovered discrepancy in isocentre co-ordinates while dubble checking print out of

treatment chart.

11/3/2016