Previous Page 852-853 / 3581 Next Page
Page Background

-10.0

-8.0

-6.0

-4.0

-2.0

0.0

2.0

4.0

6.0

8.0

10.0

-10.0

-5.0

0.0

5.0

10.0

cone beamZ

epid Z

Correlation CBCT – EPID for lung

-10.0

-8.0

-6.0

-4.0

-2.0

0.0

2.0

4.0

6.0

8.0

10.0

-10.0

-5.0

0.0

5.0

10.0

cone beamX

-10.0

-8.0

-6.0

-4.0

-2.0

0.0

2.0

4.0

6.0

8.0

10.0

-10.0

-5.0

0.0

5.0

10.0

cone beamY

epid Y

Left-Right

Sup-Inf

Ant-Post

Setup error measured with cone beam CT

EPID

EPID

EPID

Observer error in CBCT is negligible

large observer error in EPID

such as a clear under-estimation of AP shift (lung)

Slope: 0.81 .. 1.06

0.66 .. 0.88

0.30 .. 0.66 (95% CI)

Borst et al, IJROBP 2007