Quantitative Analysis Summary Report

Sample Type

Response

Final Conc

Exp Conc

Accuracy

Calibration

37598

10.1260

10.0000

101.26

Calibration

71576

21.2075

20.0000

106.04

Calibration

167953

52.6403

50.0000

105.28

Calibration

317424

101.3894

100.0000

101.39

Calibration

604434

194.9961

200.0000

97.50

Calibration

831510

269.0554

300.0000

89.69

Calibration

1000088

324.0361

400.0000

81.01

Sample

6271

0.0000

Sample

39001

10.5835

Sample

40054

10.9267

Sample

39267

10.6702

Sample

39344

10.6954

Sample

39896

10.8752

Sample

39580

10.7723

Target Compound

Riboflavin

Sample Type

Response

Final Conc

Exp Conc

Accuracy

Calibration

1200

2.0272

1.0000

202.72

Calibration

1883

2.8130

2.0000

140.65

Calibration

4348

5.6511

5.0000

113.02

Calibration

7749

9.5651

10.0000

95.65

Calibration

16473

19.6067

20.0000

98.03

Calibration

40826

47.6370

50.0000

95.27

Calibration

82575

95.6892

100.0000

95.69

Calibration

173711

200.5872

200.0000

100.29

Calibration

262535

302.8232

300.0000

100.94

Calibration

354048

408.1546

400.0000

102.04

Sample

531

1.2572

Sample

88826

102.8850

Sample

86761

100.5081

Sample

87018

100.8037

Sample

87653

101.5343

Sample

86242

99.9109

Sample

87028

100.8153

01-01-2015-WSV-VAL-P&A-1.xlsx