ESTRO IMRT course 2015

Dose Calculation in Static and Rotational IMRT

Brussels October 8th, 2015

-5-

Precise dose calculation is still a challenge:

Be aware of potential software BUGS!

Example:

depth dose curve

in water

for a central

1x1 cm field

, 6MV

results of a dose calc. algorithm (here: CC) of a widely used commercial TPS...

calculated depth dose curve,

1mm voxel size

measured depth dose curve

calculated depth dose curve,

3mm voxel size

courtesy P. Lang,

C. Heinz

(LMU Munich)

typically challenges for all dose

calculation algorithms:

small fields, irregular shaped

fields (esp. off-axis), high

energies (15MV and more)