White et al. IJROBP 68:547-554. 2007
CBCT guidance for APBI
20 patients with
CBCT after skin-mark setup
and before treatment delivery
image registraton using
ipsilateral lung
and
external contours
random and systematic error were calculated and compared
Setup errors in mm
skin-mark setup
CBCT guidance*
RL
AP
SI
RL
AP
SI
systematic
2.7 1.7 2.4
0.8
0.7 0.8
random
2.4 2.2 2.9
1.5
1.5 1.6
isotropic CTV-PTVmargin
8.8
3.6
RL: right-left, AP: anterior-posterior, SI: superior-inferior, *3 mm action level (shifts > 3 mm are executed)
skin-mark setup
for APBI patients is sufficient for
1 cm CTV-PTV
margins
CBCT guidance
may permit a reduction of CTV-PTV margin
by a factor of 2