Table of Contents Table of Contents
Previous Page  335 / 1228 Next Page
Information
Show Menu
Previous Page 335 / 1228 Next Page
Page Background

The problem …

Energ

CSDA range

1

For brachytherapy photons:

d

ll l i

h

y

[keV]

-

electrons [g/cm

2

]

[cm]

30

1.8E-03

2.5

100

1.2E-02

5.6

350

1 1E 01

8 9

secon ary e- ranges are sma re at ve to p oton

m.f.p.

e- radiative energy loss is negligible

. -

.

1000

4.4E-01

14

CPE can be assumed to exist at all points (except

close to a source or high Z materials)

Dose to a point can be approximated by collision KERMA throughout a geometry

of mm sized voxel elements

In short, to know the dose distribution one needs to know the energy distribution

of fluence Φ at all points of a geometry:

,

Ε

,

Table from: Venselaar, Baltas, Meigooni, Hoskin (Eds), Comprehensive Brachytehrapy: physical and

clinical aspects. CRC Press, Taylor & Francis, © 2013