AAPM TG-53 TPS Commissioning

Confirmation of dose model input data (

from publications

) for

each source type. The basic

literature datasets

selected for use

and comparisons should be identified.

Comparison of single point, 2-D and 3-D dose distributions with

hand calculations

for a single source, for each source type in the

source library.

Comparison of point, 2-D and 3-D dose distributions with

hand

calculations

for multiple source configurations, for at least one

source type.

Any

applicator shielding effects

included or neglected should be

explained and documented.

Verify correct behavior of dose calculations, sometimes

including

tissue multiple scattering and attenuation

, at selected distances

from the source.

Fraass, et al.,

Med. Phys