Type B uncertainty … input

Phantom geometry (for single source dosimetry)

Remember dose from scatter depends on geometry dimensions…!

or patient segmentation and source/applicator position (for patient dosimetry)

Geometry in simulation for S

K

(for single source dosimetry)

In order to comply with the definition of S

K

:

Air (point) detector at large distance from the source in vacuo

Photon emissions of energy lower than δ=5 keV (i.e. characteristic x-rays of Ti

encapsulation following photoelectric absorption of 125I photons) must be

suppressed.