Chapter 3: Calibration of Brachytherapy Sources

3.3 In-air measurement technique

3.4 Calibration using well type chambers

3.5 Calibration using solid phantoms

Measurements in solid phantoms are not suitable for low-energy sources.

3.6 Relative measurements

Readings of consecutive source deliveries can be compared and deviations larger than

3% or 5% should be investigated. Serious incidents may be identified before treatment.

ESTRO Booklet N

0

8 (2004)