Chapter 6: Calibration Using Free In-air Measurements

6.1. General

The free in-air measurement technique cannot be used for low-energy

sources due to air-kerma calibration factor uncertainties, low air-kerma rates,

and uncertainties due to air humidity. For long-lived radionuclides, e.g.,

137

Cs,

a source can be a working standard.

6.2. Formalism for reference air kerma rate

K

R

reference air kerma rate

N

K

chamber air kerma calibration factor at desired photon energy

M

U

measured charge, corrected for T, P, recombination, transit error

t

time for collecting charge

k

scat

correction for room scatter

k

n

correction factor for non-uniform electron fluence within the cavity

d

measurement distance from source center to chamber center

d

ref

reference measurement distance (i.e., 1 meter)