Chapter 5: Calibration at the SSDL and Hospital Level

5.1. Establishment of standards for photon and intravascular sources

5.1.1. Traceability in calibrations at SSDLs

The recommended detector is an appropriately calibrated well type chamber.

The preferred traceability method is to have the well type chamber calibrated

against the primary standard at the PSDL. Calibrations at an ADCL or the

IAEA Dosimetry Laboratory can serve as an alternative. This calibration

should be carried out for each radionuclide and source type to be used.

5.1.2. Traceability in calibrations at hospitals

It is recommended that for brachytherapy sources be calibrated with an

appropriately calibrated well type chamber. For traceability, the well type

chamber should be calibrated at the SSDL (or ADCL).