Precision data were generated with all SPIFAN I matrices. Accuracy (spike recovery) data were

generated with 1 fortified and 4 placebo SPIFAN I matrices.

2.

10 of the SPIFAN matrices were analyzed with immunoaffinity columns from two different

manufacturers, EASI-EXTRACT (Biopharm) and BioTeZ IAC (Biotez)

3.

Representative chromatograms provided.

-

Method Optimization:

Performance Characteristics:

Analytical Range: 0.1-300 µg/100g. SMPR = 0.1-150 µg/100g

LOQ: Estimated to be 0.1 µg/100g. SMPR <0.1 µg/100g.

Accuracy/Recovery:

Results for SRM 1849a are within the certified range.

Spike recoveries range from 95-105%. SMPR 90-110%

Precision (RSD

r

): Repeatability was not calculated. Overall RSD of all results ranged from 0.55-

3.19%. SMPR repeatability <6%.

Reproducibility (RSD

R

): NAP

System suitability: No system suitability requirements listed, but authors noted that during the SLV, r

2

was not less than 0.998.

Recommendation:

All of the data from analyses of SPIFAN matrices met the SMPR requirements. I would recommend this

method for first action status.