FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

1

Expert Review Panel for Infant Formula and Adult Nutrition

Evaluation of Method:

Bio-02

#3

Title:

Determination of biotin by liquid chromatography coupled with immunoaffinity column clean-up extraction

Author:

George Joseph, AsureQuality

Summary of Method:

The sample is dissolved, then autoclaved. After cooling, the diluted sample is centrifuged and filtered. The

filtrate is collected for clean-up and extraction using an immunoaffinity column is mounted onto a SPE

manifold. Biotin from the column is eluted with methanol and collected in a reacti-vial. The eluent is

evaporated to dryness and the sample is re-constituted in 1mL of water. The biotin in the reconstituted

sample is quantified by HPLC using a UV detector set at 200nm

Method Scope/Applicability:

Infant Formula and Adult/Pediatric Formulas

General comments about the method:

Method Clarity:

Clear directions for operation of the method, instruments. Clear equations describing how the standards

and samples are quantitated is required.

Method Safety Concerns:

None, beyond standard laboratory safety

Pros/Strengths:

Simple straight forward analysis.

Cons/Weaknesses

Proprietary components (immunoaffinity cartridge)

Supporting Data

SPIFAN Kit 1 used.

Data not calculated on RTF basis

Performance Characteristics:

Analytical Range:

0.1

to 300 µg/100g meets SMPR limits (0.1-150 µg/100g)

Bio-02 Review Forms