FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

General Comment:

Method Optimization:

No data about the use of blanks and no data about calibration curve of standards included in the report.

Performance Characteristics:

Analytical Range: data not included

SMPR 200 – 10000

µ

/100g

LOQ: data not included

SMPR ≤ 200

µ

/100g

Accuracy/Recovery: 96 – 110 %

SMPR 90-110% for Vit B3

The recovery was calculated on all SPIFAN samples and on NIST reference material SRM 1849a. The results

obtained ( 96 – 110 % ) show that the method meets de SMPR recovery for niacin.

Precision (RSD

r

): ≤ 5 %

SMPR : ≤ 5 % for Vit B3

The precision was calculated on 3 SPIFAN samples, in triplicate and the method meets de SMPR.

Product

Niacinamide

Nicotinic

Acid

1849a-1

101452

4122

1849a-2

97718

4202

1849a-3

100242

4058

Ave.

99804

4127

RSD

1.9

1.8

Child Nut. -1

83461

3909

Child Nut. -2

85461

3848

Child Nut. -3

85795

3862

Ave.

84906

3873

RSD

1.5

0.8