FOR EXPERT REVIEW PANEL USE ONLY
NOT FOR DISTRIBUTION
2
General Comment:
Method Optimization:
No data about the use of blanks and no data about calibration curve of standards included in the report.
Performance Characteristics:
Analytical Range: data not included
SMPR 200 – 10000
µ
/100g
LOQ: data not included
SMPR ≤ 200
µ
/100g
Accuracy/Recovery: 96 – 110 %
SMPR 90-110% for Vit B3
The recovery was calculated on all SPIFAN samples and on NIST reference material SRM 1849a. The results
obtained ( 96 – 110 % ) show that the method meets de SMPR recovery for niacin.
Precision (RSD
r
): ≤ 5 %
SMPR : ≤ 5 % for Vit B3
The precision was calculated on 3 SPIFAN samples, in triplicate and the method meets de SMPR.
Product
Niacinamide
Nicotinic
Acid
1849a-1
101452
4122
1849a-2
97718
4202
1849a-3
100242
4058
Ave.
99804
4127
RSD
1.9
1.8
Child Nut. -1
83461
3909
Child Nut. -2
85461
3848
Child Nut. -3
85795
3862
Ave.
84906
3873
RSD
1.5
0.8
BVit-02 Review Forms