3
Analytical Range:
Not clearly specified
LOQ:
Based on spike and recovery data a LOQ of 1.3 μg/100 g was observed for lutein and 1.5 μg/100 g for β-
carotene. SMPR is < 1 ug/100 g but since there is a concentration step in sample preparation LOQ could
probably go down if more sample is taken for evaporation
Accuracy/Recovery:
Limited data but meet SMPR:
Accuracy by spike recovery of SRM1849a at 2 levels of all-trans lutein and all-trans b-carotene –
meet SMPR for recoveries and repeatability (no reference or certified carotenoid values available
from NIST)
Precision (RSD
r
):
Limited data but meet SMPR
Reproducibility (RSD
R
):
no data available
System suitability:
there is some information on how to determine resolution between cis and trans
isomers but need to specify acceptance criteria
Recommendation:
Expand the scope to include a-carotene and lycopene
Complete validation to determine repeatability and accuracy for different samples
All SPIFAN matrices need to be analyzed
Clarify system suitability requirements
Add data about analytical range
See if LOQ could be improved to meet SMPR
Decision on recommending this method as AOAC First Action is better made after data for more SPIFAN
matrices are collected.
Carot-02 Review Forms