Table of Contents Table of Contents
Previous Page  89 / 159 Next Page
Information
Show Menu
Previous Page 89 / 159 Next Page
Page Background

FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

1

Expert Review Panel for Infant Formula and Adult Nutrition

Evaluation of Method Chol-08/Carn-07

Title: Determination of Free and Total Choline and Carnitine in Infant Formula and Adult/Pediatric

Nutritional Formula by Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS)

Author:

David Ellingston, Covance

Summary of Method:

An analytical method for the analysis of free as well as total carnitine and choline in infant formula

and adult/pediatric nutritional formula was developed with a simultaneously analysis of both the

analytes using strong cation exchange (SCX) chromatography

.

The free portion is analyzed using

water extraction, while the total extraction employs acid assisted microwave hydrolysis. The

total amount can include contribution from bound sources such as phosphatidylcholine or acetylcarnitine.

Both extraction methods employ LCMSMS analysis with electrospray

ionization (ESI). Calibration standards are included through each extraction procedure.

Method Scope/Applicability:

The analysis of free as well as total carnitine and choline in infant formula

and adult/pediatric nutritional formula

General comments about the method:

It’s an easy method to specifically analyze total as well as free choline and carnitine in the samples. The

extract ion of analytes present in free form is a simple water extraction. The extraction of the analytes in

the bound form is a relatively faster (about 1 hr) using microwave hydrolysis. The analysis is MS/MS based

and thus specific. The method employs labeled choline and carnitine as internal stds.

Method Clarity:

Good. The authors may provide a Table providing concentrations of analytes in each level of working

standard including internal std.

#2

Chol-08 Review Forms