Page Background

6/3/2017

Varian leakage: comparison between

120- and 80-leaf MLCs

Blue: 80-leaf MLC

RED: 120-leaf MLC

-20.0

+20.0

0.0

100

50

0

Relative scale (%)

Distances from central axis (cm)

2.5%

2.0%

optical density

calibration scans

Varian’s new

“Halcyon” calims

leakage of <0.4%

Elekta leakage: comparison between

MLCi, Beam Modulator and Agility

78

0,0

0,5

1,0

1,5

2,0

2,5

3,0

3,5

4,0

-75

-50

-25

0

25

50

75

Transmission (%)

Displacement (mm)

MLC Transmission

MLCi

Beam

Modulator