Machine related QA:

Small field effects

Small number of monitor units

Tongue and groove effects

Verify Leaf Positions

Output tolerance tighter

Isocenter, mechanical tolerance tighter (smaller target)

Dosimetric accuracy

Immobilization

79

ESTRO physics course 2017

IMRT quality assurance

ICRU 83

Patient specific QA A.2.3

Individual beams

No agreement on pass–fail criteria for the gamma value

Absorbed dose in Phantom

Dosimetric accuracy of 5%

Independent absorbed-dose calculation

The independent absorbed dose calculation must be able to

compute the absorbed dose in 3D so that the calculation can be

tested at a statistically relevant number of points

In vivo dosimetry

TLD, Diode, MOSFET, EPID Dosimetry

80