IMRT audit results
•
Dose determination audits with anthropomorphic phantoms show , in the
target volume, an agreement between different centres of:
–
About 3% (1 standard deviation) for 3D-CRT
–
Up to 8% (1 standard deviation) for IMRT
•
30% of clinics failed to deliver correct IMRT H&N plans (7%/4mm)
according to Ibbot et al. (IJROBP 2008), due to:
–
Incorrect output factors and percent depth dose data
–
Inadequate modeling of the penumbra at MLC leaf ends
–
Incorrect application of QA calculations or measurements
–
Inadequate QA of multileaf collimator
–
Incorrect patient positioning
–
Errors in treatment planning software
05/04/2016